Acute and long-term (2-years) clinical outcomes of the CoreValve 31 mm in large aortic annuli: A multicenter study

2017 
Abstract Introduction Little is known about the early and late performance of the 31mm CoreValve Revalving System (CRS, Medtronic Inc., Galway, Ireland). Our aim was to compare acute and long-term results of the 31mm CRS with other valve sizes. Methods Consecutive patients undergoing transcatheter aortic valve implantation (TAVI) with CRS in nine Italian centers were prospectively included and dichotomized according to prosthesis size in two different groups, as follows: 31mm and other valve sizes (i.e., 23, 26, and 29mm combined). End points were defined according to Valve Academic Research Consortium definitions. Propensity score matching was performed. Results In total, 2069 patients (n=169 [8%] in the 31mm group and n=1900 [92%] in the other valve sizes group) were included. After propensity matching, the implantation of the 31mm valve was associated with lower rates of procedural- (91.3% vs. 98.1%, p=0.030) and device-success (88.5% vs. 97.1%, p=0.016), longer procedural time (120 [80–180] min. vs. 90 [60–120] min., p Conclusions The acute performance of the 31mm CRS was worse than other valve sizes but no negative impact was observed in long-term outcomes.