Metabolomic Analysis of Glioma Cells Using Nanoflow Liquid Chromatography–Tandem Mass Spectrometry

2018 
: Mass spectrometry (MS)-based techniques have been finding utility as sensitive, high throughput metabolite analysis tools for complex biological samples. We describe here a nanoflow liquid chromatography-tandem mass spectrometry (nano-LC-MS/MS) system we developed and applied to metabolic profiling of human cells. Metabolites are extracted from cells using methanol, and filtered through a C18 StageTip to remove large particles. Metabolite samples are separated by HPLC at a flow rate of 400-500 nl/min, then analyzed in both positive and negative ion modes in an LTQ-Orbitrap MS. Metabolite identification and differential analysis are performed using commercial or open source software. Protocols outlined in this chapter describe how nano-LC-MS can be applied to investigate metabolic profiling with limited biomass amount.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    4
    Citations
    NaN
    KQI
    []