Comparison of PFA-100 testing and bleeding time for detecting platelet hypofunction and von Willebrand disease in clinical practice

2003 
The PFA-100 instrument (Platelet Function Analyzer, Dade Behring) has been reported to be superior to the bleeding time (BT) as a screening test of primary hemostasis. However evaluation of this device has been principally limited to selected populations. The study’s aim was to determine testing performance in clinical practice, by comparing the PFA-100 to the BT for the identification of von Willebrand disease (VWD) and intrinsic platelet hypofunction. From 1998-2000, PFA-100 closure time (CT) for epinephrine-collagen (EPI) and ADP-collagen (ADP) cartridges and modified Ivy BTs were performed on outpatients referred for testing for suspected or known hemorrhagic diathesis (n=346). Evaluation included assays of von Willebrand factor and platelet aggregometry in addition to platelet flow cytometry and electron microscopy when indicated. The normal distribution of PFA-100 CTs was determined using blood samples from 61 normal donors studied on 155 occasions. Results show that thirty-four patients met the diagnostic criteria for VWD and 31 patients were diagnosed with congenital or acquired intrinsic platelet hypofunction. The sensitivity of the PFA-100 for identification of VWD was significantly better (p