Determination of chlorine ions in raw milk by pulsed amperometric detection in a flow injection system

2018 
ABSTRACT Chloride ion concentration in milk was determined by pulsed amperometric detection in a flow injection system. Results showed that the Au electrode lost 3 electrons at 1.10 V and formed chloroaurate ions (AuCl 4 − ) by combining with chloride ions, after which AuCl 4 − was partly reduced to Au at 0.6 V. Based on the electrochemical process, a triple waveform with detection potential of 1.15 V, detection time of 150 ms, oxidation potential of 1.4 V, oxidation time of 550 ms, reduction potential of 0 V, and reduction time of 400 ms was applied to the Au electrode for detecting chloride ion concentration in milk. The approach is rapid and automatic and features a detection limit of 0.005 g/L. The relative standard deviation obtained by 60 repetitive injections reached 1.48% at 2 g/L of NaCl. The method developed using the Au electrode without modification was used to analyze the chloride ion concentration in raw milk without preprocessing. The method showed good agreement with potentiometric titration.