Assessment of the reliability of the motor unit size index (MUSIX) in single subject “round-robin” and multi-centre settings

2019 
Abstract Objective The motor unit size index (MUSIX) is incorporated into the motor unit number index (MUNIX). Our objective was to assess the intra-/inter-rater reliability of MUSIX in healthy volunteers across single subject “round robin” and multi-centre settings. Methods Data were obtained from i). a round-robin assessment in which 12 raters (6 with prior experience and 6 without) assessed six muscles (abductor pollicis brevis, abductor digiti minimi, biceps brachii, tibialis anterior, extensor digitorum brevis and abductor hallucis) and ii). a multi-centre study with 6 centres studying the same muscles in 66 healthy volunteers. Intra/inter-rater data were provided by 5 centres, 1 centre provided only intra-rater data. Intra/inter-rater variability was assessed using the coefficient of variation (COV), Bland-Altman plots, bias and 95% limits of agreement. Results In the round-robin assessment intra-rater COVs for MUSIX ranged from 7.8% to 28.4%. Inter-rater variability was between 7.8% and 16.2%. Prior experience did not impact on MUSIX values. In the multi-centre study MUSIX was more consistent than the MUNIX. Abductor hallucis was the least reliable muscle. Conclusions The MUSIX is a reliable neurophysiological biomarker of reinnervation. Significance MUSIX could provide insights into the pathophysiology of a range of neuromuscular disorders, providing a quantitative biomarker of reinnervation.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    18
    References
    7
    Citations
    NaN
    KQI
    []