A Comparison of AneuRx Aortic Cuff and Zenith Distal Flare Exclusion of Common Iliac Artery Ectasia for Endovascular Aneurysm Repair

2004 
Stent-grafts are ideally terminated within the common iliac artery (CIA). However, CIA ectasia may require hypogastric artery occlusion, with stent-graft extension to the external iliac artery. Alternatively, the diameter of the distal stent-graft may be increased, or flared, to allow exclusion of the abdominal aortic aneurysm. This report details the authors’ experience with this technique. Forty-one patients received bifurcated stent-grafts (BSG): 20 received an AneuRx device, and 21 received a Zenith device. CIA ectasia (diameter 15–25 mm) was treated with a distal flare of 2–4 mm greater than the CIA diameter. Patients were followed up with computed tomography scan at 1, 6, and 12 months. Statistical analysis was performed using ANOVA within groups and unpaired two-tailed t test between groups. A p value of <0.05 was considered significant. Eight of 20 patients (40%) (11 CIA) received an AneuRx device and 13/21 (62%) (17 CIA) received a Zenith device, with a distal flare. Values are (n) mean (mm) ± SE...