Determination of trace selenium in edible fungi with graphite furnace atomic absorption spectroscopy
2006
In the present article, samples were digested by a quartz high-pressure digestion pot, reducing the loss of selenium in digestion. The content of selenium in edible fungi was determined by using graphite furnace atomic absorption spectroscopy, and the results showed that when the content of selenium in edible fungi was determined by using 1% Ni(NO_3)_2 as a matrix modifier, ashing temperature of 500 ℃, and atomization temperature of 2 500 ℃, and rectifying background by deuterium light, the recovery was in the range of 92.1%-115.5%, the relative standard deviation of the method was 1.28%, and the limit of detection was 15.8 μg·L -1. The method was suitable for the determination of trace selenium in edible fungi with the advantages of being simple, rapid, sensitive, stable and accurate etc., and the results were satisfactory.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
0
Citations
NaN
KQI