Influence Of Limb Position On Assessment Of Nerve Mechanical Properties Using Shear Wave Ultrasound Elastography
2020
INTRODUCTION: Evaluation of nerve mechanical properties has the potential to improve assessment of nerve impairment. Shear wave velocity, as measured by using shear wave (SW) ultrasound elastography, is a promising indicator of nerve mechanical properties such as stiffness. However, elucidation of external factors that influence SW velocity, particularly nerve tension, is required for accurate interpretations. METHODS: Median and ulnar nerve SW velocities were measured at proximal and distal locations with limb positions that indirectly altered nerve tension. RESULTS: Shear wave velocity was greater at proximal and distal locations for limb positions that induced greater tension in the median (mean increase proximal 89.3%, distal 64%) and ulnar (mean increase proximal 91.1%, distal 37.4%) nerves. DISCUSSION: Due to the influence of nerve tension when SW ultrasound elastography is used, careful consideration must be given to limb positioning.