Application of ultrasound during electrode implantation for sacral neuromodulation in patients with neurogenic bladder secondary to spinal cord disease: a retrospective study.

2021 
PURPOSE To compare the use of intraoperative ultrasound with X-ray fluoroscopy during sacral neuromodulation lead electrode placement in patients with neurogenic bladder secondary to spinal cord disease. METHODS We reviewed the medical records of 52 patients who underwent sacral neuromodulation (SNM) lead electrode implantation under fluoroscopy or ultrasound guidance from July 2016 to July 2019. The operating time, number of electrode contacts with stimulus responses, minimum voltage that causes a stimulus response, and rate of standard lead electrode placement were used to assess the differences between the two methods. All patients were evaluated by recording bladder diaries, postvoid residual volumes before and during the testing period. Permanent SNM implantation is acceptable if symptoms improve by at least 50%. RESULTS The operating time decreased from 87.1 ± 25.19 min in the X-ray group to 68.2 ± 25.20 min (p   0.05). There was no radiation exposure during the operation in the ultrasound group. No incisional infections, hematomas, or other critical complications were reported in either groups. CONCLUSION Ultrasound can be applied to safely place lead electrode for sacral neuromodulation and leads to no radiation exposure to the patient, surgeon, and operating room staff and a shortened operating time while maintaining the same efficacy as X-ray.