Acetylene measurement using quantum cascade lasers at 14μm
2011
The monitoring of acetylene (C 2 H 2 ) concentrations is important for many chemical processes. Industrial trace gas
measurements are usually performed using gas chromatographs (GC) which have time constants of several minutes.
Optical analyzers are expected to yield faster response times with lower maintenance costs. We investigated the use of
quantum cascade laser (QCL) spectroscopy in the 14μm range for the sensitive and fast detection of C 2 H 2 . This spectral
range is favorable, as it avoids spectral interferences by other components which could be present in typical process
gases. We developed new custom DFB QCLs and characterized their spectral properties. We determined the
performance of our QCL gas analyzer setup and demonstrate a noise equivalent concentration of 10 ppb in 20 s average
time.