Artifacts of Functional Electrical Stimulation on Electromyograph
2014
The purpose of this study is to investigate different factors of the artifact in surface electromyography(EMG) signal caused by functional electrical stimulation(FES). The factors investigated include the size of stimulation electrode pads, the amplitude, frequency, and pulse width of the stimulation waveform and the detecting electrode points. We calculate the root mean square(RMS) of EMG signal to analyze the effect of these factors on the M-wave properties. The results indicate that the M-wave mainly depends on the stimulation amplitude and the distribution of detecting electrodes,but not on the other factors. This study can assist the reduction of artifact and the selection of detecting electrode points.