A spectrophotometric flow injection system for streptomycin determination in veterinary samples.

2014 
Abstract In this work a spectrophotometric flow injection analysis system for streptomycin determination in veterinary samples, is being proposed. The method is based on streptomycin alkaline hydrolysis that forms guanidine, followed by the reaction with Fe(II). The colored product has absorption peak at 520 nm. To evaluate and optimize the system parameters, chemometrics tools, such as factorial design, Pareto chart and Doelhert design, were used. The veterinary samples are diluted in water and introduced in the FIA system, therefore no sample preparation is required. The optimized system presented: linear range of 60 up to 1000 mg L −1 , limit of detection of 18 mg L −1 and sampling rate of 36 readings per hour. The precision was checked and the CV for veterinary sample readings were always less than 6.5%. The accuracy was studied by comparison with chromatographic method, thus, five samples of pharmaceutical veterinary were determined by HPLC and by the proposed method, and the results are in agreement ( t -test , p  = 0.05).
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    3
    Citations
    NaN
    KQI
    []