Angioscopic and optical coherence tomographic evaluation of neointimal coverage: 9 months after expandable polyterafluoroethylene covered stent implantation

2017 
An expandable polytetrafluoroethylene (ePTFE) covered stent is generally employed to seal coronary artery perforation. The frequency of ePTFE covered stent use is relatively low; thus, only a handful of studies have reported neointimal coverage and endothelialization inside the deployed ePTFE and clinical time course after ePTFE implantation. This case report presents a 78-year-old man treated with an ePTFE covered stent when he suffered from coronary artery perforation after the implantation of two everolimus eluting stents in the left anterior descending artery. Follow-up coronary angiography 9 months after ePTFE covered stent implantation depicted favorable stent patency. Optical coherence tomography showed thin and uneven stent strut coverage at the culprit. Angioscopy also depicted partial white-coated coverage and stent strut exposure. The outcome of this case suggested that long-term dual antiplatelet therapy should be prescribed for preventing thrombosis after ePTFE covered stent implantation.