Semi-automated liquid-liquid back-extraction in a 96-well format to decrease sample preparation time for the determination of dextromethorphan and dextrorphan in human plasma
2002
t A semi-automated, 96-well based liquid-liquid hack-extraction (LLE) procedure was developed and used for sample preparation of dextromethorphan (DEX), an active ingredient in many over-the-counter cough formulations, and dextrorphan (DOR), an active metabolite of DEX, in human plasma. The plasma extracts were analyzed by liquid chromatography-tandem mass spectrometry (LC-MS-MS). The analytes were isolated from human plasma using an initial ether extraction, followed by a back extraction from the ether into a small volume of acidified water. The acidified water isolated from the back extraction was analyzed directly by LC-MS-MS, eliminating the need for a dry down step. A liquid handling system was utilized for all aspects of liquid transfers during the LLE procedure including the transfer of samples from individual tubes into a 96-well format, preparation of standards, addition of internal standard and the addition and transfer of the extraction solvents. The semi-automated, 96-well based LLE procedure reduced sample preparation time by a factor of four versus a comparable manually performed LLE procedure.