A new method for measurement of sinoatrial conduction time.

1978 
This study describes a new method (NM) for estimation of sinoatrial conduction time (SACT), which utilizes constant atrial pacing (AP) instead of the premature atrial beats (PABs) used in the method reported in 1973 by Strauss et al. The SACTs were obtained by both methods in 20 patients. The SACT by the Strauss method (SM) was calculated as A2A3 minus A1A1. The NM consists of high right AP for a train of eight consecutive beats at rates less than or equal to 10 beats/min faster than the sinus rhythm. The interval between the last paced atrial electrogram (Ap) and the first escape atrial electrogram (A) of sinus origin (Ap-A) was measured along with several post pacing sinus cycles. The SACT by the NM was calculated as follows: SACT = Ap-A minus A1A1. The effect of AP at higher rates was also analyzed. In two patients, the SACT with the SM could not be defined, as all the A2A3 intervals were fully compensatory; with the NM the SACT was 217 and 320 msec. In the remaining 18 patients the SACT was obtainable...