Performance and cutoffs for liver fibrosis staging of a two-dimensional shear wave elastography technique

2020 
OBJECTIVES: To assess performance and cutoffs of the 2-dimensional shear wave elastography technique available on the Aplio i800 ultrasound system (Canon Medical Systems, Japan), using transient elastography as reference standard, and to assess the correlation of shear-wave-speed dispersion with liver fibrosis or steatosis. METHODS: This was a single-center cross-sectional study. The correlations between values obtained with transient elastography and 2-dimensional-shear wave elastography, and between shear-wave-speed dispersion and fibrosis or steatosis, were assessed with Pearson's r. The diagnostic performance of the 2-dimensional-shear wave elastography for staging significant fibrosis and severe fibrosis compared to transient elastography was assessed using the area under the receiver operating characteristic curve analysis. RESULTS: Three hundred sixty-seven patients (198 males and 169 females) were studied. There was a high correlation between 2-dimensional-shear wave elastography and transient elastography (r = 0.87, P 7 and > 9 kPa. Shear-wave-speed dispersion showed a high correlation with fibrosis (r = 0.85, P < 0.0001), whereas there was a very weak correlation with steatosis. CONCLUSIONS: The results of this study show that this 2-dimensional-shear wave elastography technique is accurate for staging liver fibrosis. Shear-wave-speed dispersion is highly correlated with liver fibrosis but not with steatosis.