Achieving a Safe Endotracheal Tube Cuff Pressure in the Prehospital Setting: Is It Time to Revise the Standard Cuff Inflation Practice?

2016 
ABSTRACTNumerous studies have reported unsafe endotracheal tube (ETT) cuff pressures (CP) in the prehospital environment. The purpose of this study was to identify an optimal cuff inflation volume (CIV) to achieve a safe CP (20–30 cmH2O). This observational study utilized 30 recently harvested ovine tracheae, which were warmed from refrigeration in a water bath at 85°F prior to testing. Each trachea was intubated with five different ETT sizes (6.0–8.0 mm), and each size tube was tested with six cuff inflation volumes (5–10 cc). The order of ETT size for each trachea and CIV for each size ETT was randomly pre-assigned. Data were descriptively summarized and categorized before mixed-effects logistic regression was used to determine optimal CIV. Only 113 CP measurements (12.6%, N = 900) were within the optimal range (M = 54.75 cmH2O, SD = 38.52), all of which resulted from a CIV 6 or 7 cc (61% and 39%, respectively). CIVs of 5 cc (n = 150) resulted in underinflation (<20 cmH2O) in all instances, while CIVs o...