A Novel Method for Continuous Online Glucose Monitoring in Humans: The Comparative Microdialysis Technique

2001 
The aim of this study was to prove the feasibility of continuous subcutaneous glucose monitoring in humans using the comparative microdialysis technique (CMT). The performance of the CMT was determined by comparing tissue glucose values with venous or capillary blood glucose values in healthy volunteers and type 1 diabetic subjects. The CMT is a microdialysis-based system for continuous online glucose monitoring in humans. This technique does not require calibration by the patient. Physiological saline with glucose (5.5 mM) is pumped in a stop-flow mode through a microdialysis probe inserted into the abdominal s.c. tissue. Tissue glucose concentration is calculated by comparing the dialysate and perfusate glucose concentrations. The time delay due to the measurement process is 9 min. We tested the CMT on six healthy volunteers and six type 1 diabetic patients for 24 h in our clinical setting. Comparisons were made to HemoCue analyzer (Angelholm, Sweden) capillary blood glucose measurements (healthy volunt...