A validated CFD model to predict O2 and CO2 transfer within hollow fiber membrane oxygenators

2011 
Hollow fiber oxygenators provide gas exchange to and from the blood during heart surgery or lung recovery. Minimal fiber surface area and optimal gas exchange rate may be achieved by optimization of hollow fiber shape and orientation (1). In this study, a modified CFD model is developed and validated with a specially developed micro membrane oxygenator (MicroMox). The MicroMox was designed in such a way that fiber arrangement and bundle geometry are highly reproducible and potential flow channeling is avoided, which is important for the validation. Its small size (VFluid=0.04 mL) allows the simulation of the entire bundle of 120 fibers. A non-Newtonian blood model was used as simulation fluid. Physical solubility and chemical bond of O2 and CO2 in blood was represented by the numerical model. Constant oxygen partial pressure at the pores of the fibers and a steady state flow field was used to calculate the mass transport. In order to resolve the entire MicroMox fiber bundle, the mass transport was simulat...