The Adjustable Continence Therapy System for Recurrent Female Stress Urinary Incontinence: 1-Year Results of the North America Clinical Study Group

2009 
Purpose: We determined the efficacy, safety, adjustability and technical feasibility of the adjustable continence therapy device (Uromedica, Plymouth, Minnesota) for the treatment of recurrent female stress urinary incontinence.Materials and Methods: Female patients with recurrent stress urinary incontinence were enrolled in the study and a defined set of exclusionary criteria were followed. Baseline and regular followup tests to determine eligibility, and to measure subjective and objective improvement were performed. A trocar was passed fluoroscopically and with digital vaginal guidance to the urethrovesical junction through small incisions between the labia majora and minora. The adjustable continence therapy device was delivered and the balloons were filled with isotonic contrast. The injection ports for balloon inflation were placed in a subcutaneous pocket in each labia majora. Device adjustments were performed percutaneously in the clinic postoperatively. An approved investigational device exemptio...