Finite element modeling, validation, and parametric investigations of a retinal reattachment stent

2017 
Summary A new retinal reattachment surgical procedure is based on a stent which is deployed to press the retina back in place. An eye-stent finite element model studied the strain induced by the stent on retina. FEM simulations were performed for several stent geometric configurations (number of loops, wire diameter, intraocular pressure). The FEM was validated against experiment. Parametric studies demonstrated that stents could be successfully designed so that the maximum strain would be below permanent damage strain threshold of 2%.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    18
    References
    1
    Citations
    NaN
    KQI
    []