Development and Validation of On-line Column Switching HPLC-MS/MS Method for 10 Phthalate Metabolites in Human Urine

2010 
Phthalates, such as di (2-ethylhexyl) phthalate (DEHP), dibutyl phthalate (DBP) have been proved to be teratogenics and endocrine disruptors, metabolized rapidly and excreted in the urine. In this study, a simultaneous analytical method for 10 phthalate metabolites, MnBP, MiBP, MBzP, MCHP, MEHP, MEHHP, MEOHP, MnOP, MiNP and MiDP, in human urines, based on switching system with on-line pretreatment column using HPLC-MS/MS has been developed. This method was validated according to the guideline of bioanalytical method validation of National Institute of Toxicological Research. Limits of detection range between 0.2 and 0.9 ng/ml for 10 phthalate metabolites. The calibration curves showed linearity in the range 0.997~0.999, and the results of the intra- and inter-day validations were in the range from 0.4 to 14.7% RSD and from 0.3 to 9.4% RSD, respectively. Recoveries of phthalate metabolites varied from 87.0 to 116.1%. This analytical method showed high accuracy and stable precision for all metabolites, and seems to be suitable for biomonitoring of phthalates in human urine.