A comparative analysis of internal bone remodelling concepts in a novel implant for direct skeletal attachment of limb prosthesis evaluation: A finite element analysis:

2018 
Nowadays, numerous internal bone remodelling concepts are under development, in order to estimate long-term functionality of implants by evaluating the intensity of stress-shielding effect. This effect is also analysed for the implants for direct skeletal attachment, considered as a better exoprosthesis fixation method than prosthetic sockets. Most of bone remodelling approaches are based on basic concepts, differing with certain assumptions, which may affect the accuracy of the results. This article compares commonly used internal bone remodelling concepts and evaluates the functionality of the proposed Limb Prosthesis Osseointegrated Fixation System for direct skeletal attachment of limb prosthesis in comparison with two currently available implants: the Intraosseous Transcutaneous Amputation Prosthesis and the Osseointegrated Prostheses for the Rehabilitation of Amputees. Three concepts were chosen: without and with lazy zone and with the use of quadratic formula which considers bone overloading. There...