High-performance liquid chromatographic assay for meropenem in serum
1996
Abstract High-performance liquid chromatographic procedures have been developed for the measurement of meropenem in serum. The separation was performed on an Ultrasphere XL-ODS analytical column (75×4.6 mm I.D.). The mobile phase consisted of 10.53 mmol/l ammonium acetate-acetonitrile (95:5, v/v) (pH 4). The UV detection was at 298 nm. The quantitation limit both in serum and water was 0.25 μg/ml. The method was validated in serum and aqueous solution over the concentration range 0.25–50 μg/ml. The extraction recovery from serum spiked with meropenem was 99.7±3.4%. The intra- and inter-assay coefficients of variation were below 6%. Stored at −80°C for three months at various concentrations in serum and in aqueous solution, meropenem did not reveal any appreciable degradation. After 24 h, it was also stable at 4°C in serum, aqueous solution and supernatant of extraction but not at room temperature. The stability of the drug was also confirmed in serum after repeated freezing-thawing cycles at −80°C on four consecutive days.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
9
References
58
Citations
NaN
KQI