Pulsatiles Strömungsmodell mit elastischen Gefäßen für duplexsonographische Untersuchungen

2008 
: Using ultrasound duplex technique flow phenomena in patients' circulation can be examined. For the interpretation of these examinations it is necessary to have extensive knowledge on flow influencing parameters. This can be easily obtained from simplified flow models. This article describes the components of a flow model that allows examination of ultrasonic contrast media flowing through an artificial heart and vessel mimicking tubes. The artificial heart is the drive which pumps a water glycerol cellulose mixture through the circulation in a pulsatile manner. The shape of the ventricle, the compliance of the aorta, the viscosity of the flow medium and the wall elasticity of the examination vessel were taken into account. The attenuation caused by the surrounding tissue is simulated by a variable layer of castor oil. The flow model is suitable to produce flow profiles that are very similar to physiological profiles.