A scintillator geometry suitable for very small PET gantries

2017 
In this work we are describing a novel approach to the scintillator crystal configuration as used in nuclear medicine imaging. Our design is related to the coupling in one PET module of the two separate crystal configurations used so far there: monolithic and crystal arrays. The particular design we have studied is based on a two-layer scintillator approach (hybrid) composed of a monolithic LYSO crystal (5–6 mm thickness) and a LYSO crystal array with 4–5 mm height (0.8 and 1 mm pixels). We show here the detector block performance, in terms of spatial, energy and DOI information, to be used as a module in the design of PET scanners. The design we propose allows one to achieve accurate three-dimensional spatial resolution (including DOI information) while assuring high detection efficiency at reasonable cost. Moreover, the proposed design improves the spatial response uniformity across the whole detector module, and especially at the edge region. The crystal arrays are mounted in the front and were well resolved. The monolithic crystal inserted between crystal array and the photosensor, provided measured FWHM resolution as good as 1.5–1.7 mm including the 1 mm source size. The monolithic block achieved a DOI resolution (FWHM) nearing 3 mm. We compared these results with an approach in which we use a single monolithic block with total volume equals to the hybrid approach. In general, comparable performances were obtained.