Quantification of tantalum sedimentation rates in liquid embolic agents

2018 
BackgroundEndovascular treatment (EVT) of brain arteriovenous malformations has evolved from cyanoacrylate derivatives such as N-butyl cyanoacrylate, an adhesive glue, to ethylene vinyl copolymer-based liquid embolics such as Onyx® and SQUID® dissolved in dimethyl sulfoxide. Although these agents offer several advantages, their rapidly decreasing radiopacity, as a result of the sedimentation of tantalum powder, compromises visual control during EVT. This study aims to quantify and compare tantalum sedimentation rates of several liquid embolic agents, and determine their effects on radiopacity.MethodsThe rate of sedimentation of liquid embolics Onyx 18®, SQUID 12®, and SQUID 18® was measured after preparation by single x-ray exposures for a period of 30 minutes. The signal-to-noise ratios (SNRs) of the suspension of each liquid embolic was calculated at various time points as tantalum settled out of the suspension. Precipitating Hydrophobic Injectable Liquid (PHIL®) was imaged as a control.ResultsOnyx 18® ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    18
    References
    8
    Citations
    NaN
    KQI
    []