Multicentre investigation on electrically evoked compound action potential and stapedius reflex: how do these objective measures relate to implant programming parameters?

2012 
AbstractObjectivesThe aims of this study were to collect data on electrically evoked compound action potential (eCAP) and electrically evoked stapedius reflex thresholds (eSRT) in HiResolutionTM cochlear implant (CI) users, and to explore the relationships between these objective measures and behavioural measures of comfort levels (M-levels).MethodsA prospective study on newly implanted subjects was designed. The eCAP was measured intra-operatively and at first fitting through neural response imaging (NRI), using the SoundWaveTM fitting software. The eSRT was measured intra-operatively by visual monitoring of the stapes, using both single-electrode stimulation and speech bursts (four electrodes stimulated at the same time). Measures of M-levels were performed according to standard clinical practice and collected at first fitting, 3 and 6 months of CI use.ResultsOne hundred seventeen subjects from 14 centres, all implanted unilaterally with a HiResolution CII Bionic Ear® or HiRes 90K®, were included in the...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    26
    Citations
    NaN
    KQI
    []