Plantar-to-Dorsal Compared to Dorsal-to-Plantar Screw Fixation for Proximal Chevron Osteotomy: A Biomechanical Analysis

2005 
Background: A change in screw orientation in fixing the chevron proximal first metatarsal osteotomy was noted anecdotally to improve fixation strength. The authors hypothesized that plantar-to-dorsal screw orientation would be more stable than the conventional dorsal-to-plantar screw orientation for fixation of the chevron osteotomy. The purpose of this study was to determine if the load-to-failure and stiffness of the chevron type proximal first metatarsal osteotomy stabilized using plantar-to-dorsal screw fixation were greater than with the more conventional dorsal-to-plantar screw fixation method. Methods: One foot from each of eight matched cadaver pairs was randomly assigned to one of two groups: 1) fixation with a dorsal-to-plantar lag screw or 2) fixation with a plantar-to-dorsal lag screw. A proximal chevron osteotomy was then created using standard technique and the metatarsal was fixed according to previously established method. The bone was potted in polyester resin, and the construct was fitte...