VARIABILITY OF EXTRACTION CORKSCREWS: A BIOMECHANICAL STUDY
2018
BackgroundDevices are frequently used to gain sufficient purchase in a bone so that the bone itself can be manipulated to move or rupture soft tissue attachments. During hip surgery, several different extraction corkscrews are available to remove the femoral head, each with a different screw design with no evidence to suggest which is most effective. Additionally during the use of corkscrew devices, often due to the low bone density, stripping of the screw threads out of the femoral head can occur prior to its extraction, thus requiring reinsertion. The aims of this project were to measure the primary pullout and reinsertional forces of five commercially available corkscrews.MethodsPolyurethane bone models covering a spectrum of osteoporotic to normal bone densities (0.08 gcm-3, 0.16 gcm-3 and 0.32 gcm-3) were used in axial tensile testing at two insertional depths to assess the maximum pullout force of these corkscrews (Zimmer, Stryker, Medacta, JRI, Depuy).ResultsThere are significant differences betwee...
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
0
Citations
NaN
KQI