A REPLACEMENT FOR THE KIRSCHNER WIRE: THE POTENTIAL OF A BIOABSORBABLE PIN IN PAEDIATRIC FRACTURE MANAGEMENT

2018 
Kirschner wires are commonly used in paediatric fractures, however, the requirement for removal and the possibility of pin site infection provides opportunity for the development of new techniques that eliminate these drawbacks. Bioabsorbable pins that remain in situ and allow definitive closure of skin at the time of insertion could provide such advantages. Three concurrent studies were performed to assess the viability of bioabsorbable pins across the growth plate. (1) An epidemiological study to identify Kirschner wire infection rates. (2) A mechanical assessment of a bioabsorbable pin compared to Kirschner wires in a simulated supracondylar fracture. (3) The insertion of the implants across the physis of sheep to assess effects of the bioabsorbable implant on the growth plate via macroscopic, pathohistological and micro-CT analysis. An infection rate of 8.4% was found, with a deep infection rate of 0.4%. Mechanically the pins demonstrated comparable resistance to extension forces (p=) but slightly inf...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []