Modifiable factors associated with achievement of high-volume post-dilution hemodiafiltration: results from an international study.
2015
Background: The aim was to investigate factors associated with the successful achievement of ≥21 l/session of substitution fluid volume in patients on post-dilution hemodiafiltration. Methods: 3315 patients treated in 6 European countries with the Fresenius 5008 CorDiax machine including the AutoSub Plus feature were considered. Variables that showed a relationship with convection volume were entered in a multivariable logistic regression model. Results: Mean blood flow was 379 ± 68 ml/min. Median substitution volume was 24.7 L (IQR 22.0-27.4 L). Mean filtration fraction was 28.3 ± 4.1%. 81.5% of sessions qualified as high-volume HDF (substitution volumes ≥21 L). Higher age, dialyzer surface area, blood flow and treatment time were positively associated with the achievement of ≥21 L substitution volume; higher body mass index, male gender, higher hematocrit, graft or catheter vs. fistula, and start of week vs. mid-week were negatively associated. Conclusions: Dialysis center policy in terms of blood flow, treatment time, filter size, and perhaps even hemoglobin targets plays a key role in achieving high-volume HDF. All of these are modifiable factors that can help in prescribing an optimal combination of dialyzer size, achievable blood flows, and treatment times.