Motion Artifacts in Cardiac CT

2010 
Purpose: Cardiovascular applications of CT are primarily limited by temporal resolution of the scanner. Recent development in scanner technology has greatly increased temporal resolution. We here describe a standardized method of assessing temporal properties of various CT techniques.Material and Methods: the Novacor left ventricular assist device was mounted in a water-filled circulation phantom and scanned at different pump rates with a spiral CT unit and an electron beam unit. We also evaluated the use of ECG-triggered subsecond scanning on a spiral CT unit.Results: Using the fastest conventional scanning protocol, severe motion artifacts occurred. These artifacts could not be reproduced from image to image, even if the pump rate was adjusted to scan rate (1/s). Electron beam tomography (EBT) reproducibly yielded few artifacts at 100 ms and practically no artifacts at 50 ms scanning time. Even without ECG-triggering, pump motion could be reproduced as a cine-cycle. With the ECG-triggered partial scanni...