Influence Of Dust Loading On The Alpha-particle Energy Resolution Of Continuous Air Monitors For Thin Deposits Of Radioactive Aerosols

2002 
Alpha-particle continuous air monitors must sometimes be operated in dusty environments where significant dust loading of the filter can be anticipated. It is important to understand how this dust loading affects the response of the continuous air monitors. Not only must a filter be changed if there is a reduction in airflow, but a change may be necessary if the energy resolution deteriorates and the continuous air monitor loses sensitivity and specificity for the radioactive aerosols of interest. A series of experiments were conducted to investigate alpha-particle energy resolution of continuous air monitor filters, particularly under dust loading conditions. Aerosol particles of various sizes were tagged with radon decay products to serve as surrogates for radioactive aerosols of interest such as plutonium or uranium. While the size of radioactive aerosols, filter type, and dust type affected the energy resolution, the thickness of an underlying (nonradioactive) dust layer did not show significant effect for the materials studied and a loading range of 0.01–10 mg cm−2. Our results indicate that it is possible for continuous air monitors to detect the release of radioactive aerosols with little deterioration in energy resolution under conditions of significant dust loading provided that the deposited layer of radioactive aerosols remains thin (≤ 0.1 mg cm−2).