Positive end-expiratory pressure oscillation facilitates brain vascular reactivity monitoring

2012 
The pressure reactivity index (PRx) identifies optimal cerebral perfusion pressure after traumatic brain injury. We describe a method to improve PRx precision by induced variations in arterial blood pressure (ABP) using positive end-expiratory pressure (PEEP) modulation (iPRx). Neonatal swine (n = 10) were ventilated with static PEEP and then with PEEP oscillated between 5 and 10 cmH2O at a frequency of 1/min. PRx was recorded as a moving correlation coefficient between ABP and intracranial pressure (ICP) from spontaneous ABP activity (0.05-0.003 Hz) during static PEEP. iPRx was similarly recorded with PEEP oscillation-induced ABP waves. The lower limit of autoregulation (LLA) was delineated with continuous cortical laser Doppler flux monitoring. PEEP oscillation increased autoregulation-monitoring precision. The ratios of median absolute deviations to range of possible values for the PRx and iPRx were 9.5% (8.3–13.7%) and 6.2% (4.2–8.7%), respectively (P = 0.006; median, interquartile range). The phase-a...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    27
    Citations
    NaN
    KQI
    []