A new data-driven gated PET/CT free of misregistration artifacts.

2020 
We developed a new data-driven gated (DDG) PET/CT to improve the registration of CT and DDG PET. METHODS We acquired 10 repeat PET/CT and 35 cine CT scans for the mitigation of misregistration between CT and PET data. We also derived end-expiration phase CT as DDG CT for attenuation correction of DDG PET. Radiation exposure, body mass index (BMI), scan coverage, and effective radiation dose were compared between repeat PET/CT and cine CT. Of the 35 cine CT patients, 14 (capturing 59 total tumors) were compared among average PET/CT (baseline PET attenuation correction by average CT), DDG PET (DDG PET attenuation correction by baseline CT), and DDG PET/CT (DDG PET attenuation correction by DDG CT) for registration and quantification without increasing the scan time for DDG PET. RESULTS Compared with repeat PET/CT, cine CT had significantly lower scan coverage (32.5±11.5 cm vs. 15.4±4.7 cm; p 0.1). Cine CT saved the scan time for not needing a repeat PET. The SUV ratios of average PET/CT, DDG PET, and DDG PET/CT to baseline PET/CT were 1.14±0.28, 1.28±0.20, and 1.63±0.64, respectively (p<0.0001), suggesting that the SUVmax increased consecutively from baseline PET/CT to average PET/CT, DDG PET, and DDG PET/CT. Motion correction with DDG PET had a larger impact on quantification than registration improvement with average CT did. The biggest improvement in quantification was from DDG PET/CT, in which both registration was improved and motion was mitigated. CONCLUSION Our new DDG PET/CT approach alleviates misregistration artifacts and, compared with DDG PET, improves quantification and registration. The use of cine CT in our DDG PET/CT method also reduces the effective radiation dose and scan coverage when compared with repeat CT.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    28
    References
    1
    Citations
    NaN
    KQI
    []