Concordance of Lumipulse cerebrospinal fluid t‐tau/Aβ42 ratio with amyloid PET status
2020
INTRODUCTION: Cerebrospinal fluid (CSF) biomarkers can identify individuals with Alzheimer's disease (AD) pathology (eg, amyloid plaques, neurofibrillary tangles), but defined analyte cut-points using high-throughput automated assays are necessary for general clinical use. METHODS: CSF amyloid beta42 peptide (Abeta42), t-tau, and t-tau/Abeta42 were quantified by the Lumipulse platform in two test cohorts (A/B: Eisai BAN2401-201/MISSION AD E2609-301/302, n = 138; C: Knight Alzheimer's Disease Research Center (ADRC), n = 198), and receiver operating characteristic (ROC) curve analyses defined cut-points corresponding best to amyloid determinations using positron emission tomography (PET) imaging. The best-performing cut-point was then validated as a predictor of amyloid status in an independent cohort (D: MISSION AD E2609-301/302, n = 240). RESULTS: Virtually identical t-tau/Abeta42 cut-points ( approximately 0.54) performed best in both test cohorts and with similar accuracy (areas under ROC curve [AUCs] [A/B: 0.95; C: 0.94]). The cut-point yielded an overall percent agreement with amyloid PET of 85.0% in validation cohort D. DISCUSSION: Lumipulse CSF biomarker measures with validated cut-points have clinical utility in identifying AD pathology.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
43
References
16
Citations
NaN
KQI