Flow-injection chemiluminescence determination of haemoglobin in the blood.

2013 
In this study, a sensitive and simple flow-injection chemiluminescence (CL) method was developed for the quantitative analysis of haemoglobin. The method is based on the ability of haemoglobin to enhance the CL signal generated by a H2O2–K4Fe(CN)6–fluorescein alkaline system enhanced by CdTe quantum dots. Under the optimized conditions, haemoglobin can be detected in concentration range 7.35 × 10–9–2.5 × 10–6 mol/L, with a detection limit (3σ) of 1.8 × 10–9 mol/L and a relative standard deviation (RSD; for 5 × 10–7 mol/L haemoglobin) of 2.06% (n = 11). The present CL method was successfully applied for the determination of haemoglobin in three kinds of blood samples taken from an infant, an adult man, an adult woman and two reference samples. Compared with previous reports, the CL method described in this work is simple and rapid, with high sensitivity. Copyright © 2012 John Wiley & Sons, Ltd.