Adaptive optics with combined optical coherence tomography and scanning laser ophthalmoscopy for in vivo mouse retina imaging

2018 
Optical coherence tomography (OCT) and scanning laser ophthalmoscopy (SLO) are two state-of-the-art imaging technologies commonly used to study retina. Adaptive Optics (AO) methodologies enable high-fidelity correction of ocular aberrations, resulting in improved resolution and sensitivity for both SLO and OCT systems. Here we present work integrating OCT into a previously described mouse retinal AO-SLO system, allowing simultaneous reflectance and fluorescence imaging. The new system allows simultaneous data acquisition of AO-SLO and AO-OCT, facilitating registration and comparison of data from both modalities. The system has data acquisition speed of 200 kHz A-scans/pixel, and high volumetric resolution.