First report of the point-of-care TEG: A technical validation study of the TEG-6S system

2016 
AbstractThrombelastography (TEG) measured by the TEG5000 Hemostasis Analyzer is an established but the labor-intensive method for assessing global hemostasis. The first true point-of-care TEG, the TEG6s system, uses resonance-frequency viscoelasticity measurements and a disposable multi-channel microfluidic cartridge to assess hemostasis and response to antiplatelet therapy. TEG assays (n = 5,100) were performed on the blood of healthy volunteers (n = 157) and patients undergoing coronary revascularization at three hospitals (n = 300). The results from the TEG6s were compared with the conventional TEG5000 in accordance with Clinical and Laboratory Standards Institute (CLSI) and FDA recommendations. Precision testing was conducted using blood from healthy donors, all assays were run for 5 consecutive days in duplicate using multiple operators, lots, and instruments. Reference ranges were comparable between the TEG systems.Deming regression analysis demonstrated a strong correlation between the two systems ...