Assessment of ECH stray radiation levels at the W7-X Michelson Interferometer and Profile Reflectometer

2019 
Electron Cyclotron Heating and Electron Cyclotron Current Drive are key components for heating and control in magnetically confined fusion plasmas. The high power microwaves are not always completely absorbed leading to stray radiation [1], [2]. At W7-X, the total injected microwave power can be up to 7.5 MW @140 GHz while the entire Electron Cyclotron Emission picked-up by an observer at the edge of the plasma is a fraction of a mW. In the situation of a Michelson Interferometer, the principle measurement is the entire ECE spectrum. Thus, any stray radiation is bound to enter the spectrum. In this work initial stray radiation measurements without filters at the location of two microwave receivers -the Michelson Interferometer and the Profile Reflectometer -are discussed. The data is used to dimension a notch filter to be used with the broad band Michelson Interferometer.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    2
    References
    0
    Citations
    NaN
    KQI
    []