Assessing bias in total mercury results after removing a subsample from the bottle

2016 
ABSTRACTU.S. EPA Method 1631 for total mercury (THg) analysis in water recommends that bromine monochloride (BrCl) be added to the original bottle in which the sample was collected, to draw into solution any Hg that may have adsorbed to the bottle walls. The method also allows for the removal of a subsample of water from the sample bottle for methylmercury (MeHg) analysis prior to adding BrCl. We have demonstrated that the removal of a subsample from the sample bottle prior to THg analysis can result in a positive concentration bias. The proposed mechanism for the bias is that ‘excess’ inorganic Hg, derived from the subsample that was removed from the bottle, adsorbs to the bottle walls and is then drawn into solution when BrCl is added. To test for this bias, we conducted an interlaboratory comparison study in which nine laboratories analysed water samples in fluorinated polyethylene (FLPE) bottles for THg after removing a subsample from the sample bottle, and analysed a replicate sample bottle from whic...