Attenuation Compensation in Single-Photon Emission Tomography: A Comparative Evaluation

1982 
Attenuation of photons in single-photon emission tomography (SPECT) makes three-dimensional reconstruction of unknown radioactivity distributions a mathematically intractable problem. Approaches to approximate SPECT reconstruction range from ignoring the effect of photon attenuation to incorporating assumed attenuation coefficients into an iterative reconstruction procedure. We have developed a computer-based simulation method to assess the relative effectiveness of attenuation compensation procedures. The method was used to study four procedures for myocardial SPECT using an infarct-avid radiopharmaceutical, /sup 99m/Tc stannous pyrophosphate. Reconstructions were evaluated by two criteria: overall (sum-of-squares) accuracy, and accuracy of lesion sizing. For moderate- to high-contrast studies there were no significant differences among the reconstructions by either evaluation criterion; for low contrast ratios the iterative method produced lower sum-of-squares criterion; for low contrast ratios the iterative method produced lower sum-of-squares error. We conclude that the additional expense of the iterative method is not justified under the conditions of this study. The approach used here is a convenient tool for evaluating specific SPECT reconstruction alternatives.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    13
    References
    26
    Citations
    NaN
    KQI
    []