Rapid determination of clenbuterol residues in urine by high-performance liquid chromatography with on-line automated sample processing using immunoaffinity chromatography.
1990
Abstract A liquid chromatographic column-switching system for automated sample pretreatment and determination of clenbuterol in calf urine, using an immunoaffinity precolumn with Sepharose-immobilized polyclonal antibodies against clenbuterol, is described. A second precolumn packed with C 18 -bonded silica was used for the reconcentration of desorbed clenbuterol prior to the analytical separation. Urine, after 2-fold dilution with buffer (pH 7.4), was loaded directly onto the immuno precolumn, where clenbuterol was trapped by the immobilized antibodies. This immuno precolumn has been used for more than 200 runs with standard solutions and samples. Bound analyte was desorbed with 0.01 M acetic acid and transferred, via the second precolumn, to the analytical column. The total runtime per sample was 35 min. Using a sample load of 27 ml of dilute urine and UV detection at 244 nm, the detection limit was 0.5 ng/ml. The mean recovery of clenbuterol added to a blank urine sample at the 5 ng/ml level was 82 ± 2% ( n = 5) as determined with standard solutions loaded onto the same system. Urine samples from treated animals were analysed and the clenbuterol concentrations were comparable to those obtained by high-performance liquid chromatography using solid-phase extraction for sample clean-up.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
9
References
57
Citations
NaN
KQI