HPLC-based Measurement of Glycated Hemoglobin using Dried Blood Spots Collected under Adverse Field Conditions

2018 
ABSTRACTGlycated hemoglobin (HbA1c) measured using high-performance liquid chromatography (HPLC) assays with venous blood and dried blood spots (DBS) are compared for 143 paired samples collected in Aceh, Indonesia. Relative to gold-standard venous-blood values, DBS-based values reported by the HPLC are systematically upward biased for HbA1c<8% and the fraction diabetic (HbA1c ≥ 6.5%) is overstated almost five-fold. Inspection of chromatograms from DBS assays indicates the % glycosylated calculated by the HPLC excludes part of the hemoglobin A which is misidentified as a hemoglobin variant. Taking this into account, unbiased DBS-based values are computed using data from the machine-generated chromatograms. When the DBS are collected in a clinic-like setting, under controlled humidity/temperature conditions, the recalculated values are almost identical to venous-based values. When DBS are collected under field conditions, the recalculated values are unbiased, but only about half the HbA1c values are measur...