Ultralow-dose CT with tin filtration for detection of solid and sub solid pulmonary nodules: a phantom study

2015 
Objectives:To investigate the diagnostic performance of advanced modelled iterative reconstruction (ADMIRE) to filtered back projection (FBP) when using an ultralow-dose protocol for the detection of solid and subsolid pulmonary nodules.Methods:Single-energy CT was performed at 100 kVp with tin filtration in an anthropomorphic chest phantom with solid and subsolid pulmonary nodules (2–10 mm, attenuation, 20 to −800 HU at 120 kVp). The mean volume CT dose index (CTDIvol) of the standard chest protocol was 2.2 mGy. Subsequent scans were obtained at 1/8 (0.28 mGy), 1/20 (0.10 mGy) and 1/70 (0.03 mGy) dose levels by lowering tube voltage and tube current. Images were reconstructed with FBP and ADMIRE. One reader measured image noise; two readers determined image quality and assessed nodule localization.Results:Image noise was significantly reduced using ADMIRE compared with FBP (ADMIRE at a strength level of 5 : 70.4% for 1/20; 71.6% for 1/8; p < 0.001). Interobserver agreement for image quality was excellent...