Algebraic Helmholtz inversion in planar magnetic resonance elastography

2008 
Magnetic resonance elastography (MRE) is an increasingly used noninvasive modality for diagnosing diseases using the response of soft tissue to harmonic shear waves. We present a study on the algebraic Helmholtz inversion (AHI) applied to planar MRE, demonstrating that the deduced phase speed of shear waves depends strongly on the relative orientations of actuator polarization, motion encoding direction and image plane as well as on the actuator plate size, signal-to-noise ratio and discretization of the wave image. Results from the numerical calculation of harmonic elastic waves due to different excitation directions and simulated plate sizes are compared to experiments on a gel phantom. The results suggest that correct phase speed can be obtained despite these largely uncontrollable influences, if AHI is based on out-of-plane displacements and the actuator is driven at an optimal frequency yielding an optimal pixel per wavelength resolution in the wave image. Assuming plane waves, the required number of pixels per wavelength depends only on the degree of noise.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24
    References
    110
    Citations
    NaN
    KQI
    []