TIME-RESOLVED FAST-NEUTRON IMAGING WITH A PULSE-COUNTING IMAGE INTENSIFIER

2007 
A new imaging method that combines high-efficiency fast-neutron detection with sub-ns time resolution is presented. This is achieved by exploiting the high neutron detection efficiency of a thick scintillator and the fast timing capability and flexibility of light-pulse detection with a dedicated image intensifier. The neutron converter is a plastic scintillator slab or, alternatively, a scintillating fibre screen. The scintillator is optically coupled to a pulse counting image intensifier which measures the 2-dimensional position coordinates and the Time-Of-Flight (TOF) of each detected neutron with an intrinsic time resolution of less than 1 ns. Large-area imaging devices with high count rate capability can be obtained by lateral segmentation of the optical readout channels.