Are combined spinal-epidural catheters reliable?

2000 
SUMMARY. The combined spinal-epidural (CSE) technique can rapidly relieve labor pain. However, the location of the epidural catheter is initially uncertain. In an emergency, this untested catheter may fail to provide adequate anesthesia. This study compared the efficacy of catheters placed as a part of an epidural or needle-through-needle CSE technique in laboring women. Patients requesting pain relief received either epidural (n=601) or CSE (n=1061) analgesia. All patients had a 20 gauge, closed tip multi-holed polyamide catheter (B. Braun Medical, Inc.) inserted 2‐8 cm into the epidural space. Catheters were tested to rule out intrathecal and intravascular location. Then, epidural patients received 10‐20 mL local anesthetic ∠ opioid in divided doses. CSE patients received an infusion of 0.083% bupivacaine with opioid at 10‐15 mL/h. Of the 1495 catheters that were adequately tested, those inserted as part of a CSE technique were more likely to produce bilateral sensory change and adequate analgesia than were those inserted without prior spinal analgesia (98.6% vs 98.2%, P<0.02). Stand-alone epidural catheters were more likely to produce neither sensory change nor analgesia than those inserted as part of CSE technique (1.3% vs 0.2%, P<0.02). The only catheters that failed completely and were not intravascular were stand-alone epidural catheters. In this clinical setting, catheters inserted as part of a CSE technique had a high probability of being in the epidural space and functioning appropriately. © 2000 Harcourt Publishers Ltd