Comparison of hybrid echo-planar imaging and FLASH myocardial perfusion cardiovascular MR imaging.

2005 
The purpose of this study was to compare fast single-shot gradient-echo (FLASH) and hybrid echo-planar imaging (EPI) magnetic resonance (MR) technologies regarding the relative contrast-to-noise ratio (CNR), spatiotemporal resolution, size of inducible perfusion defects, and presence of artifacts in patients with coronary artery disease (CAD). Fifteen patients with CAD underwent rest and adenosine stress gadolinium first-pass perfusion cardiovascular MR examinations with EPI and FLASH. The study was approved by the local ethics committee, and each subject gave written informed consent. The spatial resolution of the two sequences was made similar in nine patients, and the temporal resolution was made similar in six. The images were assessed for CNR, artifact, and size of inducible perfusion defects. The CNR was significantly higher with the EPI sequence, whether matched for spatial (32 vs 22 [46%], P < .001) or temporal (35 vs 23 [51%], P < .001) resolution. There was no significant difference in scoring f...